FREE SHIPPING for any order that exceeds $50!!!

BioScanner 2000™
Cholesterol Machines User Guide

Buy cholesterol machines

VIEWING THE MAIN MENU
Turning the BioScanner 2000 Cholesterol Machines On and Using the Main Menu 
Install the MEMo Chip for the test that you are going to run and depress either of the two buttons to turn the instrument on. The BioScanner 2000 cholesterol machines will perform power up diagnostics. If there are no errors, the cholesterol machines will display INSERT STRIP. If there are errors, the error message will be displayed.

If INSERT STRIP is displayed and you would like to go to the Main Menu: 

  1. Hold down both buttons at the same time. 
  2. When choices are displayed, press "NEXT" until the choice you want is displayed. 
  3. Press "ENTER" to select your choice.

Note: Pressing "ENTER" when RUN TEST is displayed takes the instrument back to the INSERT STRIP display.

The BioScanner 2000 Cholesterol Machines Main Menu:

DOING A TEST ON CHOLESTEROL MACHINES

Step One:
Insert the MEMo Chip and turn the cholesterol machines on.  Check messages (Make sure the display reads INSERT STRIP.)

Step Two:
Apply blood to strip.  (Make sure blood covers the white area completely).

Step Three:
Insert strip as far as it will go.

Step Four:
Wait--read results.

RUNNING A CONTROL TEST
A Quality Assurance Program is a standard requirement of all medical laboratories. As you use your BioScanner 2000  cholesterol machines in your home, it is recommended that you should follow the guidelines set up for clinical laboratories. Some of the standards include running a Control Test when the following circumstances occur:

In a formal Quality Assurance Program, controls are run, and results are recorded and reviewed to help spot trends and significant changes.

Note: A control is run the same way that blood is tested. Instead of a drop of blood, a drop of control is placed on the strip. (See "Doing a Test".) The expected control ranges are provided with each package of control solution.

cholesterol machines

MEMORY STORAGE
The BioScanner 2000 cholesterol machines memory can store up to 250 Glucose test results and 30 of each other test. The machines can provide the last test result, an average value for the last 10 results, and the average values of the last 14 and last 30 days' tests results.

Note: As more than 250 total Glucose tests are performed, the memory will drop the oldest value and add the new value.

To Recall Test Results in Memory:

  1. Turn your BioScanner cholesterol machines "on" by pressing either button (if not already in use). Wait for display to read INSTALL MEMO CHIP or INSERT STRIP.
  2. Hold both buttons down to enter the Main Menu.
  3. In the Main Menu, press "NEXT" until RECALL is displayed.
  4. Press "ENTER" to get to the Chemistry Menu.
  5. Press "ENTER" again to recall Chemistry name; OR To recall controls, press "NEXT", followed by "ENTER". If more than one control test has been run, select the chemistry name you would like to view.
  6. If more than one test has been run, press "NEXT" until desired chemistry name appears. Select the Chemistry name you would like to view by pressing "ENTER".
  7. Press "NEXT" to choose from LAST, LAST 10 (average), 14 DAY (average), and 30 DAY (average) and press "ENTER" to recall results. (Calculations of averages may take up to 10 seconds.)
  8. Press "NEXT" until display reads EXIT, then press "ENTER", OR Press "ENTER" to return to the Chemistry name menu.
  9. Press "NEXT" until display reads EXIT. Then press "ENTER", and repeat until RUN TEST is displayed.

CARING FOR YOUR BioScanner 200O™ cholesterol machines

TROUBLESHOOTING GUIDE
Your BioScanner 2000 cholesterol machines are built to be very reliable. Please consult the following Troubleshooting Guide to quickly fix any problem.  If a test result is displayed that you do not expect or appears "strange" to you (i.e., the result seems too high or too low), consult the following list for possible explanations.

cholesterol machines

cholesterol machines

TECHNICAL DATA

BioScanner 2000™ Cholesterol Machines Specifications:
Calibration Curve: Input from MEMo Chip in strip package
Battery Life: Approximately 300-500 tests
Operating Temperature Range: 50-104 degrees F(10-40 degrees C)
Humidity: 20-80% RH
Dimensions: Width - 2.5 in., Length - 4.25 in., Height - 0.5 in.
Weight: -3.0 oz.
Computer Compatibility: RS-232 port for download to PC

Whole Blood Glucose
Measuring Range: 20-600 mg/dL (1.1-33.3 mmoI/L)
Memory Features: 250 test results, including date and time
Expected values:
60-130 mg/dL (3.3-7.2 mmol/L)(fasting)

Blood glucose levels will change from time to time depending on the food consumed, activity level, health status, medication, stress, or exercise.

Your physician or healthcare professional will discuss target values set just for you.

A glucose level below 50 mg/dL or above 250 mg/dL may indicate a serious medical condition. If your test result should fall below 50 mg/dL or exceed 250 mg/dL, you should contact your physician or healthcare professional as soon as possible.

Total Cholesterol
Measuring Range: 100-400mg/dL (2.59-10.36mmol/L)
Memory Features: 30 test results, including date and time.

Expected values (NCEP Guidelines):

Below 200 mg/dL (5.17 mmol/L) - Desirable
200-239 mg/dL (5.17-6.18 mmol/L) - Borderline to High
Above 240 mg/dL (6.21 mmol/L) - High

Ketones
Measuring Range: 2-70 mg/dL (2.59-10.36 mmol/L)
Memory Features: 30 test results, including date and time.

Expected values:

0.5-3.0 mg/dL(5-30 mg/L) - Serum
0.2-2.8 mg/dL(2-28 mg/L) - Blood

HDL Cholesterol
Measuring Range: 25-85 mg/dL (0.65-2.20 mmol/L)
Memory Features: 30 test results, including date and time.

Expected values (NCEP Guidelines):

Above 60 mg/dL(1.55 mmol/L) - Desirable
Below 35 mg/dL(0.9 mmol/L) - High risk for Coronary Heart Disease